Comparison of refractive outcomes in post cataract surgery using measurements from immersion and contact A-scan biometry techniques
Keywords:
cataract, refractive outcomes, immersion, A-scan biometryAbstract
It is possible that different techniques used to measure axial length (AL) and anterior chamber depth (ACD) is the cause of discrepancy in refractive outcomes of cataract surgery. This study evaluated the agreement and repeatability of AL and ACD measurements using immersion and contact A-scan biometry techniques and compared the refractive outcomes from both techniques. Twenty four patients were evaluated for agreement and repeatability of AL and ACD measurements using the two different methods. The results were analyzed using Bland and Altman plots. Another 60 patients with age-related cataract were selected to compare the refractive outcomes between both methods. The IOL power was calculated using Sanders- Retzlaff- Kraff- Theoretical (SRK-T) equation. Refraction was determined between four to six weeks postoperatively and the results were analyzed using paired t-test. The results of this study showed good agreement between both techniques was noted with no significant difference detected between measurements (p > 0.05). Significant correlation was found in all parameters (AL: r = 0.99; p < 0.01, r = 0.99; p < 0.01) ACD: r = 0.91; p < 0.01, r = 0.97; p < 0.01). No significant difference in refractive outcomes of post cataract surgery was detected between the two techniques (p = 0.07). This study concludes that contact A-scan biometry and immersion techniques provide reliable results and should not be the cause of discrepancy in the refractive planned and outcome of cataract surgery.Downloads
Published
Issue
Section
License
It is a condition of publication in the Journal that authors assign copyright to the Penerbit, Universiti Kebangsaan Malaysia, using the form available on the Copyright Assignment Form page. This ensures that requests from third parties to reproduce articles are handled efficiently and consistently and will also allow the article to be as widely disseminated as possible. In assigning copyright, authors may use their own material in other publications provided that the Journal is acknowledged as the original place of publication, and Penerbit Universiti Kebangsaan Malaysia is notified in writing and in advance.
Our journal offers an open access articles, which is under the Creative Common license type : Attribution-NonCommercial-ShareAlike (CC BY-NC-SA)