CT VS MR Attenuation Correction: A Systematic Review on PET Image Quality Assessment
Keywords:
PET/CT, PET/MR, attenuation correction, image qualityAbstract
This systematic review was conducted to evaluate the image quality performance when implementing computed tomography data (CTAC) or magnetic resonance data for attenuation correction (MRAC) on positron emissiontomography (PET) images. The CTAC and MRAC were performed on image from PET/CT and PET/MR scanners, respectively. The systematic review was done based on Preferred Reporting Items for Systematic Reviews (PRISMA). Inthis study, twelve articles were included from six databases. The image performance was evaluated by overall image quality, contrast, spatial resolution, detectability, standardised uptake value (SUV) and acquisition time. Data wasshown as mean ± standard deviation and compared between CTAC and MRAC images to determine which attenuation correction method provides better image quality. Results found that PET-CTAC and PET-MRAC have similar image performance in overall image quality (p=0.93), detectabilty (p=0.84), SUVmean (p=0.84) and SUVmax (p=0.81). Meanwhile, PET-CTAC acquisition time is significantly faster than PET-MRAC by approximately two fold (p <0.05).There were no statistical analyses performed for image contrast, spatial resolution and contrast-noise-ratio due to the insufficient data. In conclusion, although PET/CT is faster than PET/MRI procedure, images yielded from CTAC and MRAC are equivalent to each other. Due to the variation of linear attenuation coefficient for each type of tissue, future review of image quality comparison can be done focusing on specific tissue or region such as soft tissue, bone and lungsto reflect the real impact of CTAC and MRAC on PET image.DOI : http://dx.doi.org/10.17576/JSKM-2023-2102-07Downloads
Published
Issue
Section
License
It is a condition of publication in the Journal that authors assign copyright to the Penerbit, Universiti Kebangsaan Malaysia, using the form available on the Copyright Assignment Form page. This ensures that requests from third parties to reproduce articles are handled efficiently and consistently and will also allow the article to be as widely disseminated as possible. In assigning copyright, authors may use their own material in other publications provided that the Journal is acknowledged as the original place of publication, and Penerbit Universiti Kebangsaan Malaysia is notified in writing and in advance.
Our journal offers an open access articles, which is under the Creative Common license type : Attribution-NonCommercial-ShareAlike (CC BY-NC-SA)