Estimation of Endotracheal Tube Cuff Pressure among Anaesthesia Providers Does Experience Matter?

Raha Abdul Rahman, Haslinda Sulong, Nadia Md Nor, Nurlia Yahya

Abstract


The endotracheal tube (ETT) cuff pressure is usually estimated and assumed to be within the appropriate range. However, the estimation of cuff pressure may be influenced by the various estimation techniques used by anaesthesia providers and their different experience level. The main aim of this study was to observe the correlation between the experience of the anaesthesia providers and their ability to estimate endotracheal cuff pressure in a correct manner. One hundred anaesthesia providers were classified into three different groups based on their months of anaesthetic experience: Group I (< 36 months), Group II (36 to 59 months) and Group III (≥ 60 months). Following intubation, the ETT cuff was inflated by the anaesthetic nurse and ETT placement verified by the anaesthesia provider. Using the pilot balloon palpation technique (PBPT), the cuff pressure was estimated to be within the appropriate pressure range, over or under-inflated. The necessary adjustments were made if needed after measurement with a VBM™ Manometer Pressure Gauge. More number of months of anaesthesia experience was not associated with a greater ability to correctly estimate the ETT cuff pressure. However, this correlation was poor (r = - 0.177). The pilot balloon palpation technique was found to be only moderately sensitive (76.5%) but poorly specific (42.9%) for correct estimation of the ETT cuff pressure, whereby PPV and NPV were 58.2% and 63.6%, respectively. The technique was 50% (95% CI 24.04 - 75.95) sensitive while correctly estimating the under-inflated cuff and 60% (95%CI 13.25-37.81) sensitive to correctly estimating the over-inflated cuff. The results depict that the experience of anaesthesia providers did not correlate with the ability to correctly estimate the ETT cuff pressure.

Keywords


Endotracheal, cuff pressures, anaesthesia, intubation, techniques.

Full Text:

PDF

References


Bunegin L, Albin MS, Smith RB. Canine tracheal blood flow after endotracheal tube cuff inflation during normotension and hypotension. Anesth Analg 1993; 76(5):1083-90.

Lomholt N. A Device for measuring the lateral wall cuff pressure of endotracheal tubes. Acta Anaesthesiol Scand 1992; 36(8)775-8.

Seegobin RD, Hasselt GL. Endotracheal cuff pressure and tracheal mucosa blood flow: Endoscopic study of effect of four large volume cuffs. Br Med J (Clin Res Ed). 1984; 288(6422):965-8.

Mandoe H, Nikolajsen L, Lintrup U, Jepsen D, Molgaard J. Sore throat after endotracheal intubation. Anesth Analg 1992; 74(6):897-900.

Hamdan AL, Sibai A, Rameh C, Kanazeh G. Short term effects of endotracheal intubation on voice. J Voice 2007; 21(6):762-8.

Hoffman R, Dahlen JR, Lipovic D, Sturmann KM. Linear correlation of endotracheal tube cuff pressure and volume. West J Emerg Med 2009; 10(3):137-9

Sengupta P, Seesler DI, Magliner P et al. Endotracheal cuff pressure in three hospitals and the volume required to produce adequate cuff

pressure. BMC Anaesthesiol 2004; 4(1): 29; 4(1):8.

OSHA.gov [Internet]. U.S Department of Labor: Occupational safety and health administration. Anesthetic gases: Guidelines in workplace exposures. [Cited 1999 July 20; revised 2000 May 18].Available from: http://www.OSHA.gov.

Noorddin Y, Raha AR, Jaafar MZ, Rozaidi SHW, Muraly S, Marlizan MY. Nitrous oxide pollution in the operating room. A comparison of two modes of ventilation. Med J Malaysia 2007; 62(2):127-9.

Vivek P, Robert JH, Allison R, Chetan B, Christine P, In-Hei H. Practicing paramedics cannot generate or estimate safe endotracheal tube cuff pressures. Prehosp Emerg Care 2007; 11(3):307-11.

Scott LS, Janet AS, Barbara RN, Richard Z. A comparison of endotracheal tube cuff pressure using estimation techniques and direct intracuff measurement. AANA J 2003; 71(6): 443-7.

Robert JH, Vivek P, In-Hei H. Experienced emergency medicine physicians cannot safely inflate or estimate endotracheal tube cuff pressure using standard techniques. Am J Emerg Med 2006; 24(2):139-43.

David L, Micheal P, Kevin C, Jay T, Nancy B. Evaluation of manual cuff palpation to confirm proper endotracheal tube depth. Prehosp Disaster Med 2008; 23(3):270-4.

Chan SM, Wong CS, Cherng CH. Determining an optimal tracheal tube cuff pressure by the feel of the pilot balloon: A training course for trainees providing airway care. Acta Anaesthesiol Taiwan 2009; 47(2):79-83.

Rose L, Redl L. Survey of cuff management practices in intensive care units in Australia and New Zealand. Am J Crit Care 2008; 17(5):428-35.

Jaber S, El Kamel M, Chanques G et al. Endotracheal tube cuff pressure in intensive care unit: the need for pressure monitoring. Intensive Care Med 2007; 33(5):917-8